Alaris® PCA module
Topic 2: Preparing an Infusion


Page 11 of 15

Glossary

Resources

Back

Next

Menu

Exit


Priming the Set with the Syringe

You can prime the PCA administration set manually using the syringe, or you can prime it using the Alaris® PCA module, if this priming option has been configured in your PCA module. Priming using the Alaris® PCA module will appear only after the syringe selection type and size and medication type have been specified, and prior to beginning an infusion. To prime the PCA administration set using the Alaris® PCA module:

  1. Press the Options key.
  2. Press the Prime Set with Syringe soft key.
  3. Ensure that the security key is in the PROGRAM position.
  4. Press and hold the Prime soft key until fluid flows and priming of administration set is complete.
  5. Release the Prime soft key when priming is complete.
  6. Press Exit to return to the main screen.