Responding to a Syringe Empty AlarmWhen the Syringe Empty alarm sounds:
If the syringe is empty, load a new syringe and restore the infusion or press the Channel Off button. If the syringe is not empty, plunger travel may be impeded. Verify that plunger movement is not impeded. It is sometimes possible for an IMS pre-filled syringe label to peel off and become lodged between the plunger and the barrel. Replace syringe if necessary. |