Alaris® Pump module
|
Page 8 of 14 |
Air-in-Line DetectionThe Alaris® Pump module is equipped with an ultrasonic component that enables the system to detect a single bolus of air and, if so configured, an accumulation of air. Your facility can select from three settings (50, 75 or 250 microliters), for each profile, the upper limit for a single bolus of air to pass without alarm. If Accumulated Air is enabled, the system will detect the presence of multiple air bubbles that are too small to be detected by the single bolus limit. The accumulated air detection limit varies according to the setting for the single bolus limit. |
|