Changing a Syringe During Infusion
If you need to change a syringe during an infusion:
- Press the PAUSE key on the Alaris® Syringe module to stop the infusion.
- Close the clamp on the extension set.
- Open the plunger grippers and syringe barrel clamp.
- Remove the syringe and separate the administration set from the syringe.
- Reattach the administration set to the new syringe and load it into
the Alaris® Syringe module.
- Press the Channel Select key.
- Select and confirm the new syringe type/size.
- Restore or reprogram the infusion.
- Open the clamp on the extension
set.
- Start infusion.
|
|