Inserting the Pressure Sensing Disc

If your hospital is using Alaris® Syringe module administration sets with pressure sensing discs, you need to insert the disc. After loading the syringe and priming the set:

  1. Gently slide the disc (cavity forward - membrane toward instrument) up into the slot in the pressure sensing disc housing.
  2. Apply firm upward pressure on the pressure sensing disc (not tubing) until the disc snaps into place.