Alaris® Syringe module
|
Page 6 of 11 |
Selecting the Syringe Type and SizeWhen you press the CHANNEL SELECT key after loading a syringe and administration set, the system will prompt you to select and confirm the syringe type and size. To do this:
Warning: Ensure the displayed syringe manufacturer and size correctly identify the installed syringe. Mismatches may cause an under-infusion or over-infusion to the patient that could result in serious injury and/or death. |
|