Priming the Set
You can prime the administration set manually, or you can prime it
with the Alaris® Syringe module. The latter option is available
only after the syringe and infusion type have been specified, and prior
to beginning an infusion. To prime the set with the Alaris® Syringe module:
- Remove the pressure sensing disc from the instrument.
- Press the OPTIONS key.
- Press the Prime Set with Syringe soft key.
- While holding the pressure sensing disc firmly between thumb and
forefinger, press and hold the PRIME soft
key until fluid flows and priming of administration set is
complete.
- Release the PRIME soft key when priming is complete.
- If necessary, expel any fluid in the pressure sensing disc by gently
massaging the membrane with a finger.
- Press EXIT to
return to the main screen.
|
|